Rapid Chloride Permeability Tester
One of the effective factors in reducing the durability of concrete is the penetration of chloride ions. Rapid Chloride Permeability Test (RCPT) allows the evaluation of chloride permeability properties in concrete. In this test method, the amount of electric current passing through a cylindrical sample with a thickness of 50 mm and a diameter of 100 mm is measured over a period of 6 hours. The potential difference of 60 volts DC is kept constant at the end of the sample, while one side of the sample is immersed in sodium chloride solution and the other side is immersed in sodium hydroxide solution. At the end, the total passing charge in the Columbus unit, which is related to the resistance of the sample to the permeability of chloride ions, is recorded as the test result. During the test, the air temperature around the sample should be maintained in the range of 20 to 25 ° C. It should be noted that the equipment of this test is used in the electrical conductivity of hardened concrete testing (ASTM C1760). RCPT Device Specifications: 1- Number of Channels, 4 Independent Channels. 2- Current Measurement (mA) 0 – 600. 3- Temperature Measurement (Celsius) 0-100. 4- PC Connection. 5-Short Circuit Protection system.
| Description |
Rapid Chloride Permeability Tester (RCPT) |
| Model |
5110151 |
| Channel |
4 Independent |
| Temp. Measurement |
0-100 Celsius |
| Current Measurement |
0-600 mA, Voltage |
| Power Supply |
110 V, 60 Hz,/ 220 V, 50 Hz |
| Display |
LCD |
| Communication Port |
USB |
| Calibration Method |
Software |
Reviews